Safety Information

Women with pencil and clipboard

Brief Summary of Safety Information for Physicians VNS Therapy System

[Depression Indication] (CE) August 2025 

The information contained in this Brief Summary for Physicians represents partial excerpts of important prescribing information taken from the physician’s manuals. (Copies of VNS Therapy physician’s and patient’s manuals are posted at www.livanova.com). The information is not intended to serve as a substitute for a complete and thorough understanding of the material presented in all of the physician’s manuals for the VNS Therapy and its component parts nor does this information represent full disclosure of all pertinent information concerning the use of this product, potential safety complications, or efficacy outcomes. Physicians should inform patients about all potential risks and adverse events discussed in the physician's manuals. This document is not intended to serve as a substitute for the complete physician's manuals. 

INTENDED USE / INDICATIONS

The VNS Therapy system is indicated for the treatment of chronic or recurrent depression for patients who are in a treatment-resistant or treatment-intolerant major depressive episode.

CONTRAINDICATIONS

Vagotomy — The VNS Therapy system cannot be used in patients with a bilateral or left cervical vagotomy.

Diathermy — Do not use shortwave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (hereafter referred to as diathermy) on patients implanted with a VNS Therapy system.

WARNINGS — ALL IMPLANTS

Use — This device is a permanent implant. It is only to be used in patients with severe depression who are unresponsive to standard psychiatric management. It should only be implanted by physicians who are trained in surgery of the carotid sheath and have received specific training in the implantation of this device. It should only be prescribed and monitored by physicians who have specific training and expertise in the management of treatment-resistant depression and the use of this device. 

Not a Cure — Physicians should warn patients that VNS Therapy has not been determined to be a cure for depression. Patients should be counseled to understand that individual results will likely vary. Beneficial results might not become evident for months. Most patients will continue to require antidepressant medications and/or electroconvulsive therapy (ECT) in addition to VNS Therapy. 

Worsening Depression / Suicidality — Patients being treated with adjunctive VNS Therapy should be observed closely for clinical worsening and suicidality, especially at the time of VNS Therapy stimulation parameter changes or drug or drug dose changes, including either increases or decreases in the stimulation parameters or concomitant treatments. Consideration should be given to changing the therapeutic regimen of VNS Therapy or concomitant treatments, including possibly discontinuing VNS Therapy or the concomitant therapy, in patients whose depression is persistently worse or whose emergent suicidality is severe, abrupt in onset, or was not part of the patient’s presenting symptoms. 

Safety and Efficacy Not Established — The safety and efficacy of the VNS Therapy system have not been established for uses outside its approved indications for use. The safety and efficacy of VNS Therapy have not been shown for people with these conditions: Acute suicidal thinking or behavior; Cardiac arrhythmias or other abnormalities; History of dysautonomias; History of previous therapeutic brain surgery or CNS injury; History of schizophrenia, schizoaffective disorder or delusional disorders; History of rapid cycling bipolar disorder; History of respiratory diseases or disorders, including dyspnea and asthma; History of ulcers (gastric, duodenal, or other); History of vasovagal syncope; Only one vagus nerve; Other concurrent forms of brain stimulation; Pre-existing hoarseness; Progressive neurological diseases other than depression; 

Dysfunctional Cardiac Conduction Systems — The safety and effectiveness of the VNS Therapy system in patients with predisposed dysfunction of cardiac conduction systems (re-entry pathway) have not been established. Evaluation by a cardiologist is recommended if the family history, patient history, or electrocardiogram suggests an abnormal cardiac conduction pathway. Serum electrolytes, magnesium, and calcium should be documented before implantation. Additionally, postoperative bradycardia can occur among patients with certain underlying cardiac arrhythmias. Post-implant electrocardiograms and Holter monitoring are recommended if clinically indicated. 

Bradycardia or Asystole During Implantation — It is important to follow recommended implantation procedures and intra-operative product tests described in the Implantation Procedure Overview. During the intra-operative System Diagnostics infrequent incidents of bradycardia and/or asystole have occurred. If asystole, severe bradycardia (heart rate < 40 bpm), or a clinically significant change in heart rate is encountered during a System Diagnostics or during initiation of stimulation, physicians should be prepared to follow guidelines consistent with Advanced Cardiac Life Support (ACLS). 

External Defibrillation or Cardioversion (electrical) — External defibrillation or cardioversion (electrical) procedures may damage the generator and can temporarily or permanently damage the nerve. Follow these recommendations to minimize the flow of current through the generator and lead system: Position defibrillation patches or paddles perpendicular to the generator and lead system, and as far from the generator as possible; Use the lowest clinically appropriate energy output (watt-seconds); Confirm generator function after any internal or external defibrillation, or cardioversion treatment. 

Magnetic Resonance Imaging (MRI) — Patients with the VNS Therapy system, or any part of the system, implanted should have MRI procedures performed only as described in the MRI Guidance instructions for use. In some cases, surgery will be required to remove the VNS Therapy system if a scan using a transmit RF body coil is needed. 

MR Unsafe Devices — The Wand, Programmer, and patient magnet are MR Unsafe devices. These devices are projectile hazards and must not be brought into the MR scanner room. 

Excessive Stimulation — Excessive stimulation is the combination of an excess duty cycle (i.e., one that occurs when ON time is greater than OFF time) and high frequency stimulation (i.e., stimulation at ≥ 50 Hz). Excessive stimulation has resulted in degenerative nerve damage in laboratory animals. 

Device Manipulation — Patients who manipulate the generator and lead through the skin (Twiddler’s Syndrome) may damage or disconnect the lead from the generator and/or possibly cause damage to the vagus nerve. Patients, parents, and caregivers should be warned against manipulating the generator and lead. 

Swallowing Difficulties — Dysphagia (difficulty swallowing) may occur with active stimulation, and aspiration may result from the increased swallowing difficulties. Patients with pre- existing swallowing difficulties and those with a history of drooling or hypersalivation are at greater risk for aspiration. Appropriate aspiration precautions should be taken for such patients. Use of the magnet to temporarily stop stimulation while eating may mitigate the risk of aspiration. 

Dyspnea or Shortness of Breath — Dyspnea (shortness of breath) may occur with active VNS Therapy. Any patient with underlying pulmonary disease or insufficiency, such as chronic obstructive pulmonary disease or asthma, may be at increased risk for dyspnea and should have their respiratory status evaluated prior to implantation and monitored following initiation of stimulation. 

Obstructive Sleep Apnea (OSA) — Patients with obstructive sleep apnea (OSA) may have an increase in apneic events during stimulation. Lowering stimulus frequency or prolonging “OFF” time may prevent exacerbation of OSA. Vagus nerve stimulation may also cause new onset sleep apnea in patients who have not previously been diagnosed with this disorder. It is recommended that patients being considered for VNS Therapy who demonstrate signs or symptoms of OSA, or who are at increased risk for developing OSA, should undergo the appropriate evaluation prior to implantation. 

Device Malfunction — Device malfunction could cause painful stimulation or direct current stimulation. Either event could cause nerve damage and other associated problems. Instruct patients, parents, and caregivers to use the magnet to stop stimulation if they suspect a malfunction, and then to contact their physician immediately for further evaluation. Prompt surgical intervention may be required if a malfunction occurs.

WARNINGS — GENERATORS

Potential Loss of Function (Model 1000 and Model 1000-D — Serial Numbers <500,000 Only) — The generator has an internal sensor that temporarily inhibits stimulation when you hold the magnet over the generator. This sensor component may fail to operate properly in a small number of Model 1000 and Model 1000-D generators, causing the following: The generator stops delivering Normal Mode therapy; The generator cannot successfully complete an on-demand diagnostic test, and an error code may display during attempts to perform diagnostic testing; If an intentional generator reset is performed while the component is in a non-functional condition, diagnostic testing may result in LOW lead impedance. For implanted Model 1000 and Model 1000- D generators, perform an on-demand System Diagnostic test at each visit to monitor the generator's status and confirm function. Advise patients to report any change in perceived clinical symptoms related to stimulation (e.g., increase in depression, changes in perception of stimulation). The diagnostic test performed during the initial interrogation to enter a session will NOT detect this event. Contact Technical Support if you observe an error code or LOW lead impedance.

PRECAUTIONS — ALL IMPLANTS

Physician Training — Physicians who prescribe should be experienced in the diagnosis and treatment of depression and should be familiar with the programming and use of the VNS Therapy system.

Use During Pregnancy — The safety and effectiveness of the VNS Therapy system have not been established for use during pregnancy.

Effects on Other Medical Devices — The VNS Therapy system may affect the operation of other implanted devices (e.g., cardiac pacemakers, implanted defibrillators). Possible effects include sensing problems and inappropriate device responses. If the patient requires concurrent implantable pacemaker, defibrillator therapy, or other types of stimulators, careful programming of each system may be necessary to optimize the patient’s benefit from each device. Furthermore, when the VNS Therapy system and another stimulator are implanted in the same patient, the two stimulators should be placed at least 10 centimeters (4 inches) apart to avoid communication interference. Users should refer to the product labeling for the concurrent device to determine if there are additional precautions that should be observed.

Device Reset — Models 103 through 1000-D and Model 8103: When the generator is reset, its stimulation output is disabled; however, all settings and device history are preserved. After a successful reset, the generator stimulation output may be re-enabled to resume operation at the previously programmed settings.

Device Reset — Model 102 and Model 102R: A reset of the device will program the device OFF (output current = 0 mA).

Device History Loss — Model 102 and Model 102R: A reset of the device causes all device history information to be lost. The device history information (e.g., programmed patient initials, implant date, device serial number) should be documented before it is reset.

PRECAUTIONS — GENERATOR AND LEAD

Battery Depletion or Drain — Model 102 and Model 102R: Do not use frequencies of 5 Hz or below for long-term stimulation.

PRECAUTIONS — RELATED TO IMPLANTATION

Vagus Nerve Placement — The VNS Therapy system is indicated for use only in stimulating the left vagus nerve in the neck area inside the carotid sheath, below where the superior and inferior cervical cardiac branches separate from the vagus nerve. The safety and efficacy of the VNS Therapy system have not been established for stimulation of the right vagus nerve or of any other nerve, muscle, or tissue.

Reversal of Lead Polarity — Reversal of lead polarity has been associated with an increased chance of bradycardia in animal studies. It is important that the electrodes are attached to the left vagus nerve in the correct orientation.

Infection Control — It is important to follow infection control procedures. Infections related to any implanted device are difficult to treat and may require that the device be explanted. The patient should be given antibiotics pre-operatively. The surgeon should ensure that all instruments are sterile prior to the operation. Careful monitoring for site infection as well as the avoidance of manipulation of the surgical site post implant should be stressed.

Lead Stabilization — The patient can use a neck brace for the first week to help ensure proper lead stabilization.

Programming After Surgery — Do not program the VNS Therapy system to an ON or periodic stimulation treatment for at least 14 days after the initial or replacement implantation. Failure to observe this precaution may result in patient discomfort or adverse events.

Laryngeal Irritation — Laryngeal irritation may result from stimulation. Patients who smoke may have an increased risk of laryngeal irritation.

PRECAUTIONS — HOSPITAL AND MEDICAL ENVIRONMENTS

VNS Therapy System Operation — Always perform device diagnostics after any of the procedures mentioned herein. Additional precautions for these procedures are described below.

Mammography — To obtain clear images, patients may need to be specially positioned for mammography procedures because of the location of the generator in the chest. Therapeutic Radiation — Therapeutic radiation may damage the generator’s circuitry. Sources of such radiation include therapeutic radiation, cobalt machines, and linear accelerators. The radiation effect is cumulative, with the extent of damage determined by the total dosage. The effects of exposure to such radiation can range from a temporary disturbance to permanent damage and may not be detectable immediately.

Electrosurgery — Use of electrosurgery [electrocautery or radio frequency (RF) ablation devices] may damage the generator.

Extracorporeal Shockwave Lithotripsy — Extracorporeal Shockwave Lithotripsy may damage the generator. If therapeutic ultrasound is required, do not position the area of the body where the generator is implanted in the water bath or in any other position that would expose it to ultrasound therapy. If that position cannot be avoided, program the generator output to 0 mA for the treatment, and then after therapy, reprogram the generator to the original parameters. 

Treatment That Involves Electrical Currents — If the patient receives medical treatment for which electric current is passed through the body (e.g., from a TENS unit), either the generator output should be set to 0 mA or the function of the generator should be monitored during the initial stages of treatment.

Therapeutic Ultrasound — Routine therapeutic ultrasound could damage the generator and may be inadvertently concentrated by the device, causing harm to the patient. Diagnostic ultrasound has no known adverse effects on the generator or lead.

Magnetic Resonance Imaging (MRI) — An MRI should not be performed using a transmit RF body coil for certain VNS Therapy device configurations or under certain specific conditions. In some cases, heating of the lead caused by the transmit RF body coil during MRI may result in serious injury. Static, gradient, and radio frequency (RF) electromagnetic fields associated with MRI may change the generator settings (i.e., reset parameters) or activate the VNS Therapy device if the Magnet Mode (epilepsy) output remains “ON”.

Receive RF Coils — Certain magnetic resonance (MR) system head coils operate in receive-only mode and require use of the transmit RF body coil. Other MR systems use a transmit/receive RF head coil. Local or surface coils may also be receive-only RF coils that require the transmit RF body coil for MRI. The use of a receive RF coil does not alter hazards of the transmit RF body coil.

Transmit RF Coils — Exposure of the VNS Therapy system to any transmit RF coil must be avoided. Do not perform MRI scans using any transmit RF coil in the defined exclusion zones.

PRECAUTIONS — HOME OCCUPATIONAL ENVIRONMENTS

Avoid Strong Electric or Magnetic Fields — Patients should exercise reasonable caution in avoiding devices that generate a strong electric or magnetic field. If a generator ceases operation while in the presence of electromagnetic interference (EMI), moving away from the source may allow it to return to its normal mode of operation.

ADVERSE EVENTS

Implant-related adverse events reported during the pivotal study in ≥ 5% of patients are listed in order of decreasing occurrence: incision pain, voice alteration, incision site reaction, device site pain, device site reaction, pharyngitis, dysphagia, hypoesthesia, dyspnea, nausea, headache, neck pain, pain, paresthesia, and cough increased.

Stimulation-related adverse events reported during the acute sham-controlled study by ≥ 5% of VNS Therapy-treated patients are (in order of decreasing occurrence): voice alteration, cough increased, dyspnea, neck pain, dysphagia, laryngismus, paresthesia, pharyngitis, nausea, and incision pain.

LivaNova USA, Inc.
100 Cyberonics Boulevard
Houston, Texas 77058, USA
Tel: +1 (281) 228-7200 (worldwide) 
Tel: +1 (800) 332-1375 (US/Canada)
Fax: +1 (281) 218-9332
LivaNova Belgium NV
Ikaroslaan 83
B-1930 Zaventem, BELGIUM
Tel: +32 2 720 95 93
Fax: +32 2 720 60 53

 

 

 

www.livanova.com

© 1998 – 2025 LivaNova, PLC, London, UK. All rights reserved.

All trademarks and trade names are the property of LivaNova or the property of LivaNova's consolidated subsidiaries and are protected under applicable intellectual property laws. Solely for convenience, LivaNova's trademarks and trade names may appear without the ® or TM symbols, but such references are not intended to indicate in any way that LivaNova will not assert, to the fullest extent under applicable law, LivaNova's rights to these trademarks and trade names. Prior permission from LivaNova is required for the use or reproduction of such intellectual property rights. The Bluetooth® word mark and logos are registered trademarks owned by BluetoothSIG and any use of such marks by LivaNova is under license.