Patient MRI Options

MRI Patient Options
MRI Patient Options
Learn more about MRI options for your patients
MRI Scans are safe with VNS Therapy™ provided specific guidelines are followed1

Learn more about MRI options for your patients

MRI with VNS Therapy™ Manual

Radiology FAQs

 Group A2Group B2Group B2
 

DemiPulse™ Model 103
AspireHC™ Model 105
AspireSR™ Model 106
SenTiva™ Model 1000
SenTiva DUO™ Model 1000-D
AND
Generator Location - in the upper left chest at or above armpit (above anterior rib 4)

DemiPulse™ Model 103
AspireHC™ Model 105
AspireSR™ Model 106
SenTiva™ Model 1000
SenTiva DUO™ Model 1000-D
AND
Generator Location - NOT in the upper left chest at or above armpit (above anterior rib 4)
Model 100C, Model 101, Pulse™ Model 102, Pulse Duo™ Model 102R, DemiPulse Duo™ Model 104
(Applicable for all implant locations)
MR Conditional
MR Conditional
YesYesYes
Static Magnetic Field Strength1.5T or 3T1.5T or 3T1.5T or 3T
Scanner TypeHorizontal field, cylindrical closed-bore clinical system for hydrogen proton imaging
Horizontal field, cylindrical closed-bore clinical system for hydrogen proton imaging
Horizontal field, cylindrical closed-bore clinical system for hydrogen proton imaging
Operating ModeNormal Operating ModeNormal Operating ModeNormal Operating Mode*
Exclusion ZoneBody coil: C7-L3
Transmit-receive head or extremity coil: C7-T8
C7-T8C7-T8
Spatial Field Gradient≤3000 Gauss/cm≤3000 Gauss/cm≤3000 Gauss/cm
Max Slew Rate200 T/m/s200 T/m/s200 T/m/s
RF CoilTransmit: Body coil or Transmit-receive head or extremity coils
Receive: No Restrictions
Transmit-receive head or extremity coilsTransmit-receive head or extremity coils
Max SARTransmit head coil: 3.2 kg
Transmit body coil: 2.0 kg
Transmit-receive head coil: 3.2 W/kgTransmit-receive head coil: 3.2 W/Kg
System ProgrammingStimulation OFF Sensing OFF*
*for select models with AutoStim mode
 Optional device features OFF (Model 1000 and Model 1000-D only)
Stimulation OFF Sensing OFF*
*for select models with AutoStim mode
 Optional device features OFF (Model 1000 and Model 1000-D only)
Stimulation OFF
Exposure TimeTransmit head or extremity coil: No restrictions
Transmit body coil: ≤15 minutes of active scan time within a 30 minute window
Transmit-receive head or extremity coil: No restrictionsTransmit-receive head or extremity coil: No restrictions
Additional RestrictionsTransmit head or extremity coil: none
Transmit body coil: Circularized Polarized mode only
NoneNone

*Normal Operating Mode: Transmit Head Coil: 3.2 W/Kg, Transmit Body Coil: 2.0 W/Kg, Transmit/Receive head coil: 3.2 W/Kg2

NOTE: MRI scans using a body coil have not been approved in all geographies.
Always review the most current labelling prior to performing an MRI scan.2

For MRI Safety information, refer to MRI Instructions For Use.